
SmartReader Plus Data Logger
44
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
1.8
SmartReader Plus 5 Thermocouple Logger
The SmartReader Plus 5 Thermocouple Logger can monitor and record temperatures from J,
K, T, S or E type thermocouples.
Figure 5-1: SmartReader Plus 5
1.8.1 Description
The SmartReader Plus 5 has two external thermocouple temperature channels and one
internal thermistor channel.
For information on temperature ranges and resolutions applicable for the different types of
thermocouples you want to use, refer to the Specifications section.
1.8.2 Setup
The internal thermistor temperature channel for the SmartReader Plus 5 is used to simulate a
reference junction for the thermocouples - it must always be enabled.
see SmartReader Plus 5 Specifications
see SmartReader Plus 5 Specifications
1.8.3 Use
1. Connect a J, K, T, S or E type thermocouple to one or more channels. Make sure that
you observe polarity when making connections. To do this, connect a thermocouple's
negative wire (usually red) to the "-" terminal of the channel you want to use. Connect the
Comentarios a estos manuales