the HVL measurement.
The test protocols
of the following two sections
describe methods to measure
the HVL to
determine
compliance with minimum filtration requirements.
Since
the
HVL
is a function of the actual high voltage applied
to the x-ray tube, the protocols
assume the unit's
kV
calibration is correct.
If such is not the case, the
measured HVL may
lead to
an erroneous
conclusion
12