SmartReader Data Logger
17
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
notes packed with each RH probe set -- see Figure 2-2.
Figure 2-2: Connecting the EH-020A Temperature and RH Probe
The EH-020A is applicable for measurements between -20 and 40°C (-4 and 104°F) and 10 to
90% RH.
NOTE: The EH-020A probe is compatible with loggers that have serial numbers
above 20,000 only.
The accurate measurement of temperature and humidity depends primarily on the information
supplied by a particular sensor. SmartReader 2 data loggers use temperature and humidity
sensors with rugged qualities designed for minimal or no maintenance (under normal
operating conditions).
The SmartReader 2's on-board thermistor is chemically stable and not significantly affected by
aging. It will typically drift less than 0.1°C over a period of several years. The relative humidity
sensor, although subject to accuracy degradation when exposed to contaminants and/or
extreme environmental conditions, will drift typically less than 1% per year (under clean
conditions). You should check your RH sensor periodically and, if necessary, recalibrate or
replace it.
1.4.4 Troubleshooting
If you are getting what appear to be wrong readings, consider the following before recalibrating
the sensors. Sensors sample the atmosphere (or medium) only in the immediate vicinity of
the sensors themselves. The physical state of the atmosphere and its degree of uniformity
and turbulence will limit the validity of a measurement at some distance from the sensor. This
becomes especially apparent for the measurement of ambient temperature and relative
humidity.