ACR Electronics SmartReader Plus Manual de usuario Pagina 39

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 103
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 38
SmartReader Data Logger
31
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
Type J Thermocouple Inputs
Thermocouple
Equation
Range
Resolution
1 (Narrow)
49
-20 to 190°C
(0 to 370°F)
1.4°C (2.5°F)
2 (Wide)
50
-50 to 600°C
(-55 to 1100°F)
5°C (9°F)
Type K Thermocouple Inputs
Thermocouple
Equation
Range
Resolution
1 (Narrow)
51
-25 to 230°C
(-10 to 440°F)
1.7°C (3°F)
2 (Wide)
52
-100 to 900°C
(-145 to 1650°F)
6.7°C (12°F)
Type S Thermocouple Inputs
Thermocouple
Equation
Range
Resolution
1 (Narrow)
73
0 to 960°C
(32 to 1750°F)
7°C (12.6°F)
2 (Wide)
74
0 to 1760°C
(32 to 3200°F)
23°C (41.5°F)
Type T Thermocouple Inputs
Thermocouple
Equation
Range
Resolution
1 (Narrow)
53
-35 to 200°C
(-30 to 390°F)
1.7°C (3°F)
2 (Wide)
54
-200 to 400°C
(-325 to 750°F)
6.7°C (12°F)
Specifications are subject to change without notice.
1.7.4 Calibration
Low, Mid and High Calibration Values
The Low calibration value compensates for offset while Mid compensates for span. These
values are set at the factory using an accurate millivolt reference. The following procedure
may be used if you want to calibrate each channel to a specific piece of thermocouple wire:
1. Change the equation number of the channel you want to calibrate to 38, and replace any
previous Low, Mid and High calibration values with zero (0). Make sure you have
Vista de pagina 38