ACR Electronics SmartReader Plus Manual de usuario Pagina 80

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 103
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 79
SmartReader Data Loggers
72
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
2. MgCl2 x 6H2O, 150gm (Fisher #M-33)
3. NaCl, 50gm (Fisher #S-271)
· De-ionized or distilled water, 75ml for each solution
Prepare for the saturated salt test by following these steps:
1. Use the powder funnel to carefully place the salt into the center of the jar bottom.
2. Slowly pour the water onto the mound of salt, using the long stem funnel, taking care not
to splash any water or salt onto the inner walls of the jar.
3. Gently swirl the saturated solution. If the salt dissolves entirely, with no crystals
remaining on the bottom, the solution is not saturated and more salt must be added.
There must be crystals remaining on the bottom but they must not be exposed above the
water level.
4. Put a nail through the jar lid and bend into a small hook. Tape the head of the nail (or
use rubber cement) to create an airtight seal of the opening. The logger with RH sensor
(or EH-020A) can be inserted into the jar by first hanging it on the nail loop then screwing
on the lid to the jar.
5. The saturated salt solutions should be kept in a well insulated environment to minimize
any temperature changes. The equilibrium RH over saturated salt solutions is only
slightly temperature dependent. However, for close calibration work to 5% RH or better, it
is important to keep the temperature of the salt solution and the air space above it
identical. To ensure this, the temperature of the lab should be constant.
To perform the saturated salt test, follow these steps:
1. Setup your relative humidity logger with a sample rate of 20 seconds and equation 38 on
the RH channel(s). Make sure you have selected Normalized Byte data units.
2. Place the jars into the box with 1 to 2 inches of Styrofoam chips on the bottom, then fill all
remaining air spaces between the jars and the interior spaces of the box with the chips.
3. Cover the box with cardboard in which 6cm (2.5") diameter holes have been cut to allow
access to the tops of the jars for insertion of the logger or RH module.
4. Place the logger with sensor or EH-020A module into the jar containing the LiCl solution
(RH = 11%) overnight to eliminate any hysteresis effects.
5. Quickly insert the logger (or module) into the next solution by swapping the lids.
Transfers should always be made in the direction of higher RH conditions. The jar
should be well sealed. Allow at least two hours to get within three percent of the
equilibrium RH. Repeat for next solution.
Salt
RH
MgCl2
33%
NaCl
75%
6. Call up the graph on your computer using TrendReader software. The profile you see
should reflect the test you have just performed.
7. Zoom in on the section that relates to the 33% RH level in the test chamber. Use the
Vista de pagina 79
1 2 ... 75 76 77 78 79 80 81 82 83 84 85 ... 102 103

Comentarios a estos manuales

Sin comentarios